- Advanced second-generation CCD full-spectrum spectral manufacturing technology (digital technology replaces the old bulky and bulky photomultiplier tube analog technology), and the channel can be realized without limitation;
- Introduce European technology, synchronize with international spectrometer technology, and be a domestic manufacturer capable of producing CCD full-spectrum technology spectrometers;
- The use of blown-type constant temperature light chamber is stable, reliable and maintenance-free, and the design of small double light chambers is no longer used to solve the technical problems of long and short wave elements ;
- There is no cost for channel changes and additions within the matrix ;
- It is convenient to upgrade multi-substrate without changing and adding hardware
- Excellent data stability, the analysis of the same sample in different time periods can obtain good data consistency;
- Small size, light weight, easy to move and install;
- High integration, high reliability, high stability;
- Power saving, material saving, and energy consumption are only 50% of ordinary spectrometers.
Technical Parameters
Spectrometer Optical System
- Paschen-Runge structure, Rowland circle optical system
- Grating focal length: 350mm
- Wavelength range: 160-510nm
- Detector: Multiple high-performance linear array CCDs
- Light chamber temperature: automatic control Constant temperature: 34ºC±0.5ºC
- Pixel resolution: 20pm
- Grating line: 24001/mm
- Primary spectral line dispersion rate: 1.2nm/mm
- External incident window for easy cleaning and replacement
- Excellent drift correction - thanks to automatic peak finding (accurate to 0.1 pixel)
- Excellent UV sensitivity
digital plasma generator
- Full digital plasma spark light source technology
- High efficiency thanks to compact design and semiconductor control technology
- High energy pre-combustion technology (HEPS)
- The excitation parameters can be customized and set by the user according to actual needs through the user interface
- Frequency 100-1000Hz
- Current 1-80A
Optimized sample stage
- Open sample stage design to meet the requirements of large sample testing
- 4mm sample stage analysis gap
- "Spray electrode" technology easily handles small samples and samples with complex geometries
- Low gas consumption, no standby flow during standby
- Custom Anisotropic Sample Adapters
- There is no need to change the spark table for different substrates
- Optimized impurity discharge system
- Robust copper sample stage for excellent heat dissipation
Computer and Readout System
data output
- Data is stored in various data formats
- Can be connected to various standard printing equipment
- Data can be sent remotely via the Internet
Software system
- Windows 10 operating system
- English spectrometer operating software
- Optional Data Maintenance Program
- Height: 73 (L)*67(W)*34(H)
- Weight: 30 kg
Power requirements:
- AC220V/50Hz
- Maximum power 450 w (host 420W, computer 30W)
- Average power 300 W (host 270W, computer 30W)
- Standby power 40
- 16 A slow blow fuse
H2 requirements
- Pure hydrogen gas, 99.999%; hydrogen pressure reducing valve (specially equipped with pressure reducing valve)
- Air pressure>4MPa: outlet instrument air pressure 0.5MPa