DRPL material cold surface temperature tester
Overview
The instrument uses the internationally popular back temperature measurement method, that is, constant pressure is applied to the sample, a constant high temperature hot plate is pressed on the front of the sample, and the temperature change process of the back of the sample is measured to achieve the thermal insulation performance of the analyzed material at high temperature. Equipped with computer to realize automatic detection. It can be used to test refractory materials, insulation materials, aerogel boards, vacuum glass, etc. Widely used in colleges and universities, scientific research institutions, quality inspection, factories and mines.
Technical indicators
MODEL |
DRPL |
Sample range |
120*120*(5~50)mm (customers can customize) |
Test hot surface temperature |
1000ºC, accuracy 0.1ºC |
Test sample |
solid materials and powdery materials |
Pressure loading method |
the computer is used to clamp the sample accurately; |
pressure range |
10000N, 20000N, 50000N; (customer optional) |
Thickness measurement method |
automatic or manual measurement |
thickness measurement range |
0-50mm, resolution 1μm |
Hot plate heating method |
silicon carbon rod heating; |
Number of sample groups |
1~2 pieces |
Composed |
temperature measurement system, heater, hot plate, cold plate, servo motor loading system and other components. Place temperature measuring thermocouples in the hot plate and the cold plate respectively. |
Testing method |
The whole process of the test adopts computer to automatically test and save the curve and test data. |
Voltage, frequency, power |
AC220V±10%; 50~60Hz; 2000W |