Customization: | Available |
---|---|
After-sales Service: | One Year Warranty |
Application: | Laboratory Apparatus |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
circuit type | Inductive feedback self-excited oscillation circuit, fully automatic ignition, coaxial cable output matching and tuning, power feedback, and closed-loop automatic control. | working frequency | 40.68MHZ ± 0.05% |
frequency stability | <0.1% | ||
Output Power | 800-1500W | ||
output power stability | ≤ 0.3% | ||
Electromagnetic Field Leakage Intensity | from the fuselage30cm | ||
Electric field strength E | < 1 V/m | ||
Magnetic field strength H | <0.2A |
Output working coil inner diameter | 25mm, 3 turns | torch | Triple Concentric Quartz Torch |
Coaxial sprayer outer diameter | 6mm | Double cylinder mist outer diameter | 35mm |
Argon Flow Meter Specifications and Carrier Gas Pressure Gauge Specifications | Plasma gas flow meter ( 100-1000 ) L/h ; (1.6 -16L/min) |
||
Auxiliary air flow meter ( 10-100 ) L/h ; (0.16 -1.66L/min) |
|||
Carrier gas flow meter ( 10-100 ) L/h ; (0.16 -1.66L/min) |
|||
Carrier gas regulator valve ( 0 -0.6MPa ) |
Photomultiplier tube negative high voltage | 200-1000V Stability <0.05% |
Photomultiplier tube current measurement range | 10-12 -10-4A |
Signal acquisition is V/F exchange | 1mV corresponds to 100Hz |
light path | Czerny-Turner | focal length | 1000mm |
Grating Specifications | Ion etched holographic grating | Groove Density | 3600 lines /mm |
resolution | ≤ 0.008nm | Line area | 80 ×110mm |
Scan wavelength range | 3600 lines /mm | wavelength range | 195-800nm |
lens | ф 30 1 : 1 imaging |
entrance slit | 20µm |
Transmitter Specifications | 78 * 105* 16 mm | exit slit | 20 µm |
Stepping motor drive minimum step | 0.0005nm |