Customization: | Available |
---|---|
Certification: | CE, ISO |
Distilled: | 100~3000 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product Structure
The product consists of 5 evaporator, 5 pre-heater, 5 vapor-liquid separator, 2 condenser, feed pumps, filters, control valves, flow meter, frame, piping and controller.
1. Evaporator
Using tube falling film evaporation technology,to ensure Reverse Osmosis Wate(Pure water)uniform distribution in the inner wall of each evaporation tube to form a water film evaporation.
The effect of pure steam for the next evaporator to provide heat, its complete evaporation, liquefied form distilled water.
2. Vapor-liquid separator
Using steam and water split four separation techniques(Steam and water centrifugation, underneath rotary vane clapboard separation, wire mesh capture water droplet separation, upper rotary vane clapboard separation) to ensure distilled water heat source and endotoxin content lower.
3. Preheater
Using U-tube exchange heat, high heat exchange efficiency, good mechanical performance, long service life.
4. Condenser
Unique double condenser design, special made multiway reciprocating structure, heat transfer fully, no need special cooling water, distilled water can reach the required temperature
5. Piping
Pipeline design has no storage, no dead ends, low equipment failure rate, pipeline to connect using new type sanitary clamp connection both inside and outside mirror polish, neat appearance, operate quickly and conveniently.
6. Configuration
Adjusting valve,Automatic compensation valve,throttle valve,conductivity meter, temperature meter selected domestic and overseas high-tech products, Beautiful appearance, complete functions
7. Water Source
The equipment required the water source must be pure water or reverse osmosis water, any natural water or city water is infeasible.
Advantages
1. Saving Steam
Steam savings, efficient use of coal resources, saving rate above 20%.
2. Saving Water
No need additional cooling water, save your precious Pure water (Reverse Osmosis Water) resources.
3. Output water High quality
The produced water quality conductivity < 0.6us/cm, (GB <= 2us/cm); bacterial endotoxin <0.125Eu/ m, (GB <= 0.252Eu / m)
4. Output water fast
After open the machine 5 minutes, immediately produced qualified distilled water.
5. Distilled water quality Automatic control
Intelligent Instruments continuous digital display the outlet conductivity of current distilled water, and can set the value of qualified conductivity.
Automatic switching emission unqualified distilled water.
Usage
The equipment apply to large, medium and small pharmaceutical companies to produce the Medical Pharmaceutical preparation water, injection water and so on, all levels of hospital preparation room, or chemical research institutes.
Technical parameters
Model | WD-IA | |||||||||
100-4 | 200-4 | 100-5 | 200-5 | 300-5 | 400-5 | 500-5 | 1000-5 | 2000-5 | 3000-5 | |
Distilled water Yield (L/H) |
100 | 200 | 100 | 200 | 300 | 400 | 500 | 1000 | 2000 | 3000 |
Steam Pressure (Mpa) |
0.2 | 0.2 | 0.2 | 0.2 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 |
Steam Consumption (Kg/H) |
35 | 50 | 30 | 45 | 60 | 80 | 100 | 200 | 450 | 680 |
Pure water Consumption (ROWater) (L/H) |
130 | 250 | 110 | 220 | 330 | 440 | 550 | 1100 | 2200 | 3200 |
Pure water Conductivity Requirement s (us/cm) |
10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 |
Distilled water Conductivity( us/cm) |
<1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m |
Multi-stage Pump Power (kw) |
0.75 | 0.75 | 0.75 | 0.75 | 0.75 | 0.75 | 0.75 | 1.1 | 1.5 | 2.2 |
Dimensions (mm)(L*W*H) |
1000 | 1200 | 1200 | 1400 | 700 | 700 | 700 | 850 | 1000 | 1200 |
500 | 600 | 500 | 600 | 1500 | 1500 | 1500 | 1850 | 2200 | 2400 | |
1600 | 2000 | 1600 | 2000 | 2450 | 2450 | 2450 | 2500 | 2800 | 2800 |
Model | WD-IA | |||||||||
100-4 | 200-4 | 100-5 | 200-5 | 300-5 | 400-5 | 500-5 | 1000-5 | 2000-5 | 3000-5 | |
Distilled water Yield (L/H) |
100 | 200 | 100 | 200 | 300 | 400 | 500 | 1000 | 2000 | 3000 |
Steam Pressure (Mpa) |
0.2 | 0.2 | 0.2 | 0.2 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 | 0.3 |
Steam Consumption (Kg/H) |
35 | 50 | 30 | 45 | 60 | 80 | 100 | 200 | 450 | 680 |
Pure water Consumption (ROWater) (L/H) |
130 | 250 | 110 | 220 | 330 | 440 | 550 | 1100 | 2200 | 3200 |
Pure water Conductivity Requirement s (us/cm) |
10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 |
Distilled water Conductivity( us/cm) |
<1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m | <1us/c m |
Multi-stage Pump Power (kw) |
0.75 | 0.75 | 0.75 | 0.75 | 0.75 | 0.75 | 0.75 | 1.1 | 1.5 | 2.2 |
Dimensions (mm)(L*W*H) |
1000 | 1200 | 1200 | 1400 | 700 | 700 | 700 | 850 | 1000 | 1200 |
500 | 600 | 500 | 600 | 1500 | 1500 | 1500 | 1850 | 2200 | 2400 | |
1600 | 2000 | 1600 | 2000 | 2450 | 2450 | 2450 | 2500 | 2800 | 2800 |