Model NO.
DRC-6FT
Condition
New
After-sales Service
One Year Free Parts
Transport Package
Plywood Packing
Trademark
Xiangyi
Origin
China
Production Capacity
100 PCS/Month

Product Description

Drug Tablet Dissolution Tester Solid Preparation, Capsule

Instrument structure and features:
The water tank is arranged in a single row. It can be installed with 6 baskets or pulp poles and can be installed with 6 vessels.
Dissolution cup and water tank have no gap design to ensure automatic center positioning. Eliminate system errors.
Electronic temperature sensor can display and monitor the temperature of each point water bath.
The time system consisting of a new architecture ensures smooth and accurate timing of the instrument.
The human-machine interface is friendly. The instrument adopts a large-screen color LCD screen. Users can enjoy intelligent menu-style operation according to the screen prompts. Each operation process gives Chinese character prompts and quantitative indicators.
With self-test function and automatic protection function, it can give a variety of fault alarm signals.
The installation position of the water bath can be adjusted and locked, and the vertical axis of the pulp bar, the rotating basket and the dissolution cup is automatically centered.
The pulp rod, basket and basket made of high quality imported 316L stainless steel material meet the requirements of Chinese Pharmacopoeia.
The instrument provides "program control test" function to execute 12 sets of preset sampling program combinations, suitable for routine repeatability test.
The instrument provides "basic test" function to change the test parameters at any time during the test, which is convenient for research tests.
The instrument provides "automatic warm-up" function to start preheating bath at unattended time, saving manpower and time.

Applicable standards:
Food and Drug Administration, "Guidelines for Mechanical Validation of Drug Dissolution Apparatus".
"Pharmacopoeia of the People's Republic of China" 2015 edition.
Industry standard "Drug dissolution tester" JB/T20076-2005.

Main Specifications:            
  • Level: inclination ≤0.5° (measured in two vertical directions after installation)
  • Basket paddle (axis) verticality: ≤0.5° off vertical
  • Dissolution cup verticality: ≤1° deviation from the vertical direction
  • Coaxiality between the vessel and the basket (paddle) axis: Offset axis ≤1.0mm
  • Basket (paddle) shaft swing: swing ≤ 0.3mm
  • Basket swing: swing ≤0.5mm
  • Depth of basket (paddle): 25±2mm
  • Basket (paddle) shaft speed: ±4%mm
  • Medium temperature in each vessel: 37oC ± 0.3oC
  • Vibration: ≤0.1mi1
  • Speed range: 25-250 rpm
  • Speed resolution: 1 rev / min
  • Water bath temperature range: room temperature -45oC
  • Temperature resolution: 0.1oC
  • Temperature control accuracy: ≤ ± 0.3oC
  • Number of sampling cycles: 1-12 different sampling periods can be preset.
  • Sampling cycle time: 1-99 hours per cycle optional, 12 sampling cycles can be counted down in stages, or continuous countdown.
  • Large screen Color LCD display, intelligent menu operation.
  • A removable electronic temperature sensor monitors the temperature at each point of the water bath.
  • Standard: 6 pulp 6 cups
  • Power maximum power: 1200W
  • With self-test and automatic protection, it can give a variety of fault alarm prompts and secondary overheat protection.
  • Dimensions: length * width * height 910mm * 400mm * 510mm
Dimensions length * width * height 910mm * 400mm * 510mm
Power maximum power 1200W
Standard 6 pulp 6 cups
Temperature control accuracy ≤ ± 0.3oC

The following is a set of control interfaces for reference:
Drug Tablet Dissolution Tester Solid Preparation, Capsule
Program control parameter setting interface:
Drug Tablet Dissolution Tester Solid Preparation, Capsule

Program control test real-time interface

Drug Tablet Dissolution Tester Solid Preparation, Capsule


Automatic warm-up monitoring interface

Drug Tablet Dissolution Tester Solid Preparation, Capsule


 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier
Product Groups
More