Customization: | Available |
---|---|
Type: | Static Laser |
Sample Feeding: | Dry Dispersion |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Measuring range | 0.02µm ~ 2,000µm |
Tank | Stainless steel Vol.500ml |
Accuracy | High measurement precision |
Measure time | <2 min |
Error range | D50 <±1% (National Certified Reference Material) |
Light source | Laser diode 632.8nm 3mw >20000h |
Light path | Parallel light path design comply with ISO13320 and ISO13321 |
Detector | Multi-element detectors in 80 grades are arranged in front, back and side positions, detection angel up to 165 degree |
Sampling Method | Wet |
Dispersion | Integrated ultrasonic 60W, ultrasonic, agitating, circulating time adjustable of etc. |
Agent | Tap water, distilled water or other specified agent |
Sample Qty. | mg to g |
Data process | Mie scattering theory and special analysis software, RS232 serial port communication connection |
Result output | Table and curve of particle size distribution, D10, D50, D90, D97, averaged particle size and specific surface area. |
Test method | Automatically finished by software, after adding sample into the tank, the computer starts dispersing and agitating as well as ultrasonic; the suspension is circulated to the testing zone. After test the cleaning is also fully automatic. |
Power | AC220V/110V ± 10% 50/60Hz |
Environment | Clean, temperature 0-40ºC |
Dimension | W840×D370×H460 mm 70kg |